Testing the Parameter Signals
Testing the Parameter Signals
52
The parameter test tests the signal path to and from the transducer
sockets, but not the transducers themselves.
1. Switch on the monitor, the recorder and the telemetry receiver.
2. Connect the appropriate transducer to each socket on the
transmitter.
3. Standing within view of the monitor, press the Nurse Call Button
on the transmitter and switch on the transmitter simultaneously.
The test runs while the Nurse Call is pressed.
Chapter 9 - Maintenance and Performance Assurance