Prerequisites
Solving Common Problems
Problem
FMP printed on
Series 50 A or 50 IP
trace but cannot be
switched off.
Chapter 5 - Fetal Movement Profile
For complete details of FMP annotation, refer to the Instructions for Use
provided with your Series 50 fetal monitor.
To use FMP during ambulant monitoring, you require a Series 50 fetal
monitor with:
•
FMP
•
a compatible software revision and telemetry interface
It is not recommended to use FMP during ambulant monitoring, as any
movement of the ultrasound transducer whilst the patient is ambulating
may be recorded as fetal movement.
Connecting the telemetry system to a monitor automatically disables
FMP. Disconnecting re-enables FMP. If you require FMP during
ambulant monitoring, switch it on using the function key on the monitor
or the barcode method.
Possible Causes
Software revision not
compatible with FMP from
Series 50 T Fetal Telemetry
System.
Fetal Movement Profile
Order software upgrade for the
fetal monitor.
See your fetal monitor's Service
Manual.
5
Solutions
35