8040A - Philips 50 T Series Instructions For Use Manual

Fetal telemetry system
Hide thumbs Also See for 50 T Series:
Table of Contents

Advertisement

Error Messages

8040A

Message
Display
Err 16
US Toco
8041A
Message
Display
Signal
Indicator
indicator
Panel
lamps
flashing
62
Cause
Transducers are
connected to the front
panel of the monitor.
Cause
Invalid telemetry mode.
Incorrect transducer
connected into
transmitter. (Only
Ultrasound and Toco
transducers can be used).
Transducers are
connected to the front
panel of the monitor.
Solution
Disconnect the transducers from
the monitor or switch off the
telemetry receiver.
Solution
Check the cable from the telemetry
receiver and, if necessary, replace it.
Check that the transducer is
compatible with Series 50T Fetal
Telemetry System.
Disconnect the transducers from
the monitor or switch off the
telemetry receiver.
Chapter 10 - Troubleshooting