Error Messages; Series 50 Family - Philips 50 T Series Instructions For Use Manual

Fetal telemetry system
Hide thumbs Also See for 50 T Series:
Table of Contents

Advertisement

Error Messages

Series 50 Family

Message
Display
Err 9
US Toco
Err 14
US Toco
Err 16
US Toco
Chapter 10 - Troubleshooting
The following error messages are directly related to telemetry and appear
on the fetal monitor. Refer to the Instructions for Use provided with your
monitor for error messages not related to telemetry monitoring.
Cause
Invalid telemetry mode.
Incorrect transducer
connected to transmitter.
Transducers are
connected to the front
panel of the monitor.
Solution
Check the cable from the telemetry
receiver and, if necessary, replace it.
Check that the transducer is
compatible with Series 50T Fetal
Telemetry System.
Disconnect the transducers from
the monitor or switch off the
telemetry receiver.
Error Messages
61