Initial Operation - HP E1343A User Manual

Relay multiplexer
Table of Contents

Advertisement

Initial Operation

1. HP-IB is Hewlett-Packard's implementation of IEEE Std 488.1-1984
Chapter 1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Use the following program example to verify initial multiplexer operation
by closing a channel and querying channel closure. The example first resets
the switchbox and then closes channel 02 of a single multiplexer module
(card number 1) in the switchbox. The program next queries the channel
closure state. A returned "1" shows that the command to close the channel
has been sent to the switchbox. A returned "0" shows that the command to
close the channel has not been sent to the switchbox.
The computer used in the example is an HP Series 200/300 computer with
HP BASIC as the program language. The computer interfaces to the
mainframe using the Hewlett-Packard Interface Bus (HP-IB)
interface select code is 7, the HP-IB primary address is 09, and the HP-IB
secondary address is 14. Refer to the HP 75000 Series B Installation and
Getting Started Guide for addressing information.
Example: Reset the switchbox and close channel 02
10 OUTPUT 70914;"*RST"
20 OUTPUT 70914;"CLOS (@102)"
30 OUTPUT 70914;"CLOS? (@102)"
40 ENTER 70914;Value
50 PRINT Value
60 END
1
. The HP-IB
Opens all channels.
Close channel 02.
Query channel 02 state.
Enter results into Value.
Display result.