Using the
Switchbox
Configuration
Verifying Correct
Logical Address
Settings
Chapter 2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Use this configuration when:
1. Simultaneously closing multiple channels on multiple modules.
2. Other multimeters/modules cannot control the multiplexer modules.
To use this configuration, set the first multiplexer module's logical address
to a multiple of 8. Set the next multiplexer module's logical address to the
next address value, the third module to the next address value, and so on.
The following shows the logical address settings of a switchbox with the
first multiplexer module address set to 112.
Logical Address
112
113
114
To program the switchbox, use the programming information in this manual.
To verify that the modules have the correct logical address setting, query the
HP E1300A/E1301A Mainframe or HP E1405B/E1406A Command
Module as follows:
HP E1301A Mainframe - From the "Select a Instrument" prompt of the
front panel, select the following menus:
SYSTEM
CONFIG?
LADDS
HP E1300A Mainframe, HP E1405B, or HP E1406A Command Module
with an RS-232 Terminal - From the "Select an Instrument" prompt, select
the following menus:
SYSTEM
CONFIG?
LADDS
If no front panel or RS-232 terminal is available, send the following SCPI
query command to the mainframe/command module:
VXI:CONF:DLAD?
The returning string contains all logical addresses in the system.
Configuring the Relay Multiplexer Modules
Module
HP E1345A/E1347A
HP E1345A/E1347A
HP E1345A/E1347A
Channels
100-115
200-215
300-315