Source - HP E1343A User Manual

Relay multiplexer
Table of Contents

Advertisement

:SOURce

Parameters
Comments
Example
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
TRIGger:SOURce BUS|EXTernal|HOLD|IMMediate
to advance the channel list during scanning.
Parameter Name
BUS
EXTernal
HOLD
IMMediate
Enabling the Trigger Source: The
the trigger source. The
source.
Using the
Command: You can use
TRIG
scan when
TRIGger:SOURce BUS
Using External Trigger Inputs: With
only one switchbox at a time can use the external trigger input at the
HP E1300/E1301 Mainframe "Event In" port. The trigger input is assigned to
the first switchbox requesting the external trigger source.
Assigning External Trigger: A switchbox assigned with
remains assigned to that source until it receives a command to change the
source to
BUS
,
HOLD
trigger source is available to another switchbox (with a
command). Another switchbox cannot receive an external trigger
EXTernal
source if the source is already assigned, or an error is generated.
Using Bus Triggers: To trigger the switchbox with
selected, use the IEEE 488.2 common command
Execute Trigger (
GET
"Trig Out" Port Shared by Switchboxes: See the
Related Commands:
*RST Condition:
TRIGger:SOURce IMMediate
Scanning Using External Triggers
In the following example, the trigger input is applied to the HP E1300/E1301
Mainframe "Event In" port.
TRIG:SOUR EXT
SCAN (@100:115)
INIT
trigger externally
Parameter Type
discrete
discrete
discrete
discrete
TRIGger:SOURce
INITiate[:IMMediate]
command enables the trigger
TRIGger[:IMMediate]
or
TRIGger:SOURce HOLD
TRIGger:SOURce EXTernal
, or
IMMediate
. When the source is changed, the external
).
,
[ROUTe:]SCAN
TRIGger
Sets trigger source to external.
Sets channel list.
Starts scanning cycle.
Advances channel list to next channel.
Relay Multiplexer Command Reference
specifies the trigger source
Range of Values
*TRG command
Event In BNC port
Hold triggering
Continuous triggering
command only selects
to advance the
is selected.
selected,
TRIG:SOUR EXT
TRIGger:SOURce
TRIGger:SOURce BUS
or the HP-IB Group
*TRG
command.
OUTPut
,
ABORt
69