Parameter Types. The following table contains explanations and examples of
parameter types you might see later in this chapter.
Parameter
Type
Numeric
Accepts all commonly used decimal representations of numbers
including optional signs, decimal points, and scientific notation.
123, 123E2, -123, -1.23E2, .123, 1.23E-2, 1.23000E-01. Special
cases include MIN, MAX, and INF.
Boolean
Represents a single binary condition that is either true or false.
ON, OFF, 1, 0.
Discrete
Selects from a finite number of values. These parameters use
mnemonics to represent each valid setting.
An example is the TRIGger:SOURce <
where
Optional Parameters. Parameters shown within square brackets ([ ]) are optional
parameters. (Note that the brackets are not part of the command and are not sent to
the instrument.) If you do not specify a value for an optional parameter, the
instrument chooses a default value. For example, consider the
ARM:COUNt? [<MIN|MAX>]
specifying a parameter, the present
MIN
parameter, the command returns the minimum count available. If you send the
MAX
parameter, the command returns the maximum count available. Be sure to
place a space between the command and the parameter.
Linking IEEE 488.2 Common Commands with SCPI Commands. Use a
semicolon between the commands. For example:
or
*RST;OUTP ON
Linking Multiple SCPI Commands. Use both a semicolon and a colon between the
commands. For example:
ARM COUN 1;:TRIG:SOUR EXT
This section describes the Standard Commands for Programmable Instruments
(SCPI) commands for the multiplexer modules. Commands are listed alphabetically
by subsystem and also within each subsystem.
Explanations and Examples
source
can be BUS, EXT, HOLD, or IMM.
command. If you send the command without
ARM:COUNt
value is returned. If you send the
TRIG:SOUR HOLD;*TRG
Relay Multiplexer Command Reference
source
> command
49