Initiate; Continuous - HP E1343A User Manual

Relay multiplexer
Table of Contents

Advertisement

Subsystem Syntax

:CONTinuous

Parameters
Comments
Example
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
subsystem selects continuous scanning cycles and starts the scanning

INITiate

cycle.
INITiate
:CONTinuous <
mode
:CONTinuous?
[:IMMediate]
INITiate:CONTinuous <
mode
the switchbox.
Parameter Name
mode
<
>
Continuous Scanning Operation: Continuous scanning is enabled with the
INITiate:CONTinuous ON
INITiate[:IMMediate]
trigger from a trigger source selected by the
advances the scan through the channel list. A trigger at the end of the channel
list closes the first channel in the list and the scan cycle repeats.
Non-Continuous Scanning Operation: Non-Continuous scanning is enabled
with the
INITiate:CONTinuous OFF
Sending the
INITiate[:IMMediate]
channel list. Each trigger from a trigger source selected by the
TRIGger:SOURce
command advances the scan through the channel list. A
trigger at the end of the channel list opens the last channel in the list and the
scanning cycle stops.
Stopping Continuous Scans: See the
Related Commands:
*RST Condition:
INITiate:CONTinuous OFF
Enabling Continuous Scans
INIT:CONT ON
SCAN (@100:115)
INIT
>
>
enables or disables continuous scanning cycles for
Parameter Type
boolean
or
INITiate:CONTinuous 1
command closes the first channel in the channel list. Each
TRIGger:SOURce
or
INITiate:CONTinuous 0
command closes the first channel in the
command.
ABORt
,
,
ABORt
ARM:COUNt
TRIGger
Enables continuous scanning.
Sets channel list.
Starts scanning cycle.
Relay Multiplexer Command Reference
INITiate
Range of Values
0|1|ON|OFF
command. Sending the
command
command.
,
TRIGger:SOURce
55

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1344aE1345aE1347a

Table of Contents