INITiate

Subsystem Syntax

:CONTinuous

Parameters
Comments
60
HP E1351A/53A 16-Channel FET Multiplexer Command Reference
The
subsystem selects continuous scanning cycles and starts the
INITiate
scanning cycle.
INITiate
:CONTinuous <
mode
:CONTinuous?
[:IMMediate]
INITiate:CONTinuous <
cycles for the switchbox or scanning voltmeter.
Parameter Name
Parameter Type
mode
Continuous Scanning Operation: Continuous scanning is enabled
with the
INITiate:CONTinuous ON
command. Sending the
first channel in the channel list. Each trigger from a trigger source
specified by the
TRIGger:SOURce
through the channel list. A trigger at the end of the channel list
closes the first channel in the channel list and the scan cycle repeats.
Continuous scanning can be stopped by the
non-downloaded scans, and by a system reset or a power off for
downloaded scan lists.
Scanning a Set Number of Cycles: The
or
INITiate:CONTinuous 0
number of times specified by the
the
INITiate[:IMMediate]
channel list. Each trigger from a trigger source selected by the
TRIGger:SOURce
list. A trigger at the end of the channel list opens the last channel in
the channel list, and if the number of cycles has not reached the
parameter, returns to the beginning of the scan list and
ARM:COUNt
closes the first channel. When the number of cycles equals the
parameter, the scanning cycle stops. Note that
ARM:COUNt
ARM:COUNt
does not apply to downloaded scan lists.
Stopping Continuous Scans: Continuous scanning can be stopped
by the
command for non-downloaded scans, and by a system
ABORt
reset or a power off for downloaded scan lists.
Related Commands:
TRIGger:SOURce
*RST Condition:
>
mode
>
enables or disables continuous scanning
Boolean
or
INITiate:CONTinuous 1
INITiate[:IMMediate]
command advances the scan
command goes through the scan list the
ARM:COUNt
command closes the first channel in the
command advances the scan through the channel
ABORt
,
ARM:COUNt
INITiate:CONTinuous OFF
Range of Values
0 | 1 | ON | OFF
command closes the
ABORt
command for
INITiate:CONTinuous OFF
command. Sending
,
*RST
,
TRIGger
,