OUTPut
[ROUTe:]
Appendix B
is equivalent to writing a one to the TRG INT in the
INITiate[:IMMediate]
Status/Control Register. The TRG INT sets itself back to zero after pulsing
the FET. If you have a multiple-module configuration and a downloaded
scan list, the module with the first channel on the scan list must be the last
one to receive the TRG INT. The scan cycle will start as soon as the card
with a valid channel receives TRG INT.
The
OUTPut[:STATe]
register-based programming.
[ROUTe:]CLOSe <
channel_list
DIR in the Status/Control Register is enabled.
D0 to D3 is set to a channel number in the Direct Channel
Configuration Register.
[ROUTe:]OPEN <
channel_list
in either the Scan Channel Configuration Register or the Direct Control
Register (whichever one closed the channel). It can also be implemented by
toggling the DIR bit in the Status/Control Register.
For
channel_list
[ROUTe:]SCAN <
in each module in a multiple-module configuration. The channel is set valid
if it is in that module, and invalid in the other modules. The channel list is
downloaded in the following way:
Set CLR SCN in Scan Control Register to one, then back to zero, to
clear the previous scan list.
DIR in the Status/Control Register is set to false.
Channel number is set in D0 to D3 in the Scan Channel
Configuration Register.
The example below shows how to set a scan list for channels
(@100,101,200,201):
Write to the Scan Channel Configuration Register on card 1:
VLD* true (0), 0000 to D3-D0
VLD* true (0), 0001 to D3-D0
VLD* false (1), XXXX to D3-D0
VLD* false (1), XXXX to D3-D0
Write to the Scan Channel Configuration Register on card 2:
VLD* false (1), XXXX to D3-D0
VLD* false (1), XXXX to D3-D0
VLD* true (0), 0000 to D3-D0
VLD* true (0), 0001 to D3-D0
HP E1351A/53A Register-Based Programming 91
command does not have an equivalent in
> is equivalent to the following:
>
can be implemented by setting VLD* false
, the complete scan list must be loaded
>
Sets channel 100.
Sets channel 101.
Reserves position in scan list for
channel 200.
Reserves position in scan list for
channel 201.
Reserves position in scan list for
channel 100.
Reserves position in scan list for
channel 101.
Sets channel 200.
Sets channel 201.