Chapter 3.Using The Hp E1351A/53A Fet Multiplexer Modules; Using This Chapter; Selecting Channels - HP E1351A User Manual

Fet multiplexer
Table of Contents

Advertisement

Using the HP E1351A/53A FET Multiplexer

Using This Chapter

Selecting Channels

Chapter 3
This chapter uses some typical examples to show how to use the
16-Channel FET and 16-Channel Thermocouple FET Multiplexer Modules.
Refer to Chapter 4, "Understanding the HP E1351A/53A FET
Multiplexers" for further information. This chapter contains the following
sections:
Selecting Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 29
Multiplexer Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 31
Connecting Switchbox Channels to Direct
Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 32
Connecting Switchbox Channels to Tree Terminals for
Making Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . Page 34
Scanning a Range of Switchbox Channels . . . . . . . . . . . . . . Page 36
Downloading a Scan List . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 39
Scanning a Switchbox without a Downloaded
Scan List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 40
Scanning a Switchbox with a Downloaded
Scan List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 41
Scanning Voltmeter Configuration with
HP E1326B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 42
Measuring Temperature Using Thermocouples
(HP E1353A Module only) . . . . . . . . . . . . . . . . . . . . . . Page 43
Individual channels within a module or multiple-module instrument are
selected with the
SCAN <
<
channel_list
>
commands.
opens/closes the specified channel. The
creates a scanning list which can be downloaded into RAM on each module.
The
command closes the first channel on the scan list, and the
INITiate
selected trigger mode advances the rest of the channels.
The FET Multiplexer Module uses break-before-make logic so that a closed
channel is opened before the next channel is closed. Only one channel can
be closed at any given time on the FET multiplexer. The only exception is
in
SCAN:MODE FRES
(4-wire resistance measurements). You can,
however, have a channel closed on two different modules in the same
instrument.
Using the HP E1351A/53A FET Multiplexer Modules
channel_list
or
>
OPEN
OPEN
and
CLOSe <
channel_list
SCAN <
channel_list
Chapter 3
Modules
and
CLOSe
>
immediately
>
command
29

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1353a

Table of Contents