Advertisement

ABORt

Subsystem Syntax
Comments
Example
56
HP E1351A/53A 16-Channel FET Multiplexer Command Reference
The
subsystem stops a scan in progress when the scan is enabled via
ABORt
the interface, and the trigger modes are
TRIGger:SOURce HOLD
ABORt
Operation. Resets the trigger system and places triggers in
ABORt
idle. The scan is stopped without resetting the scan conditions. The
scan conditions stay in the current configuration.
Stopping Scans Enabled from Interface. When a scan is enabled
from the interface, use an interface
E1301 front panel "Reset Instr" or "Clear Instr" key to stop the scan.
When the scan is enabled from the interface and the trigger source is
TRIGger:SOURce BUS
the HP E1301 front panel "Reset Instr "or "Clear Instr "keys to stop
the scan.
Stopping Scans Enabled from Front Panel. When a scan is
enabled from the HP E1301 front panel execute
interface or use the front panel "Reset Instr "or "Clear Instr" keys to
stop the scan.
Related Commands.
TRIGger
Stopping a Scan with
ABORt
TRIG:SOUR BUS
INIT:CONT ON
SCAN (@100:115)
INIT
.
.
.
ABOR
TRIGger:SOURce BUS
.
CLEAR
or
TRIGger:SOURce HOLD
,
ARM
INITiate:CONTinuous
Bus is trigger source.
Set continuous scanning.
Sets channel list.
Starts scanning cycle.
Aborts scan in progress.
or
command or the HP
, use
ABORt
or
over the
*RST
,
[ROUTe:]SCAN
Chapter 5
,