System Setup
Function
Hookup Advisor
Prompt level
Pre-acquisition
1
If Hookup Advisor is enabled, this option is overridden by Hookup Advisor.

Basic System

On the System Setup menu, use the arrow pad to scroll to Basic System until it is highlighted and
press the Enter key. The Basic System window opens
100
Description
Select Yes to enable the Hookup Advisor option, which monitors
the quality of resting ECG. The default is Yes.
Select the level at which the system acknowledges poor signal
quality. See
"Hookup Advisor" on page 68
setting up this option. The default is Red.
Select Yes to begin acquiring ECG data as soon as the system is
in Resting ECG mode. The system does not wait until the user
presses the ECG key before it starts acquiring ECG data. The latest
10 seconds of ECG data is ready for analysis when Pre-acquisition
is turned on. The default is No.
MAC™ 600
for information on
2047426-001K