System Errors - GE MAC 600 Operator's Manual

Resting ecg analysis system
Table of Contents

Advertisement

Troubleshooting

System Errors

The following table identifies some potential errors that may occur while you are operating the
system, the possible causes, and a recommended action to resolve the error.
If performing the recommended actions does not resolve the problem, contact authorized service
personnel.
Problem
appears on the screen.
Amber LED on the key pad
flashes intermittently.
appears on the screen.
The writer printhead is too
hot appears on the screen
The system does not power up
when operating from battery
power
138
Cause
No battery is installed in the
system.
The battery charge is low.
The writer door is open.
The Writer mechanism has
heated up due to heavy use.
The battery is fully discharged.
MAC™ 600
Solution
Install a battery and connect
the system to an AC wall outlet
to charge the battery
Connect the system to an
AC wall outlet to charge the
battery.
Close the writer door.
Turn off the system and power
on after 3 to 4 minutes. If
problem recurs with normal
use, inform service.
Connect the system to an
AC wall outlet to charge the
battery.
2047426-001K