3.
Apply the electrodes to the prepared area.
WARNING
WARNING
4.
Verify the leads are all connected and working properly.
NOTE:
Applying Resting Electrodes
CAUTION
PROPER LEADWIRE CONNECTION Improper connection will cause inaccuracies in the ECG. For
multi-link , trace each individual leadwire in the patient cable module to the colored connector
and then to the proper electrode to ensure that it is matched to the correct label location
Standard 12 Lead Placement
To acquire a standard 12 lead ECG, use the placement shown in the following illustrations.
2047426-001K
:
SHOCK HAZARD Ensure that conductive parts of the electrodes or lead wires do not
come in contact with other conductive parts.
This would cancel the protection provided by the isolated signal input.
:
CONDUCTIVE PARTS Keep the conductive parts of lead electrodes and associated parts
away from other conductive parts, including earth.
Use only electrodes and contact agents recommended by GE Healthcare. The signal
quality on the lead-check screen will not be indicated until the RL/N electrode has
been applied. When RA/ R becomes disconnected, the system will display Limb Lead
Disconnected.
:
MAC™ 600
Preparing the Patient
57