This window will display the options available. An asterisk (*) appears next to each option which
is currently activated on the system. Use the following instructions to activate an option on your
system,
1.
At the Option Code field, type the 12-digit option activation code and press the Enter key.
If you typed the code for an option which has been purchased for the system, an asterisk
will appear next to that option in the list.
2.
Repeat step 1 for each option to be activated.
3.
Highlight Return and press Enter to return to the Basic System menu.

Miscellaneous Setup

On the Basic Setup menu, highlight Miscellaneous Setup and press the Enter key. The
Miscellaneous Setup window opens.
The Miscellaneous Setup window setup options are defined in the following table:
Function
Buzzer
Information line
Cart number
Site number
2047426-001K
Description
Select On to turn on the system's buzzer. Select Off to turn off the
system's buzzer. The default is On.
Select Yes to enable the help information line on the screen. The
default is Yes.
Type a number that uniquely identifies this system.
Type a number from 1-32 to identify where the data will be
stored in the MUSE system. The Site number used must match
the site number on the MUSE system to which the system is
communicating.
MAC™ 600
System Setup
105