You can change the mode. Changes to the mode take effect when the
transceiver is disconnected from TeleMon. The following settings will be
used:
Mode Set at TeleMon
Continuous
5-minute
1-minute
Manual
When the transceiver is sourcing the display of patient data on an MP5 bedside
monitor:
The SpO2 measurement mode is always Continuous.
To turn SpO
monitoring off, disconnect the sensor cable from the transceiver.
2
SpO
enters a power-down mode after the cable is disconnected from the
2
transceiver, thereby conserving battery life.
It is important to disconnect the sensor from the transceiver in order to conserve
battery life. Unplugging the sensor from an adapter cable that is connected to
the transceiver does NOT provide SpO
SpO
should also be turned off at the Information Center. If the transceiver is
2
configured for Continuous SpO
without turning SpO
2
Silencing this technical alarm will turn the SpO
Mode when Transceiver
Continuous
Continuous
Continuous
Spot Check
power-down mode.
2
measurement and the sensor is removed
2
off, a SpO2T NO SENSOR technical alarm will result.
is Disconnected
measurement off.
2