Standby Mode - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Turning the Transceiver On/Off
Standby
Mode
Initiating
Standby
4-4
Basic Operation
To restart monitoring, insert batteries (if needed), attach leads to the patient,
insert the SpO
sensor cable in the SpO
2
to verify association with the Information Center.
This feature must be enabled by your service provider.
Standby mode is used to temporarily suspend monitoring at the Information
Center, for example, when a patient goes out of access point range or is
discharged.
Standby does not power down the transceiver, so if the standby period is
prolonged, you should remove the batteries.
Step
Action
1
Do not remove the leads until after the patient is in Standby. This
will avoid an "ECG Leads Off" technical alarm, as well as
reminders, if configured.
2
In the Patient Window, select the
3
Select the duration of the standby period (depending on your
system's configuration: Infinite, 4 Hours, 3 Hours, 2 Hours, 1 Hour,
30 Minutes, 20 Minutes, or 10 Minutes).
If the patient will be discharged, select "Infinite" as the
Note—
standby duration.
sensor port and press the Check button
2
button.
Standby