Philips IntelliVue TRx Instructions For Use Manual page 188

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Cleaning and Sterilization
11-8
Maintenance, Cleaning & Troubleshooting
Step
Action
9
Process the equipment to be processed as follows:
Pressure: 11 +/- 1 psig (established in the preceding step).
Time: 2-3 hours
Temperature: 54.4 +/- 2.8
10
Extract the gas mixture from the sterilizer as follows:
Warning
Warning
Comply with OSHA standards
room, but vent only outdoors or to a suitable, evacuated
container, depending upon state, provincial, or country environ-
mental regulations. (If the mixture is captured, it can be sepa-
rated commercially and the component gases re-used.)
*
See "References" on page 11-9.
a. Pump the gas mixture out of the chamber until you ob-
tain a vacuum of -26 inHg +/- 1 (-12.77 psig +/-.49), re-
turning the mixture to a suitable evacuated container.
b. Return the sterilizer chamber to ambient pressure by in-
troducing air that has been bacterially filtered.
11
Air-wash the chamber and material as follows:
a. Apply -26 inHg +/- 1 (-12.77 psig +/- .49) vacuum to the
chamber and processed material again, to remove residu-
al EO. The vacuum pump must be vented to the out-
doors.
b. Return the sterilized chamber to ambient pressure by in-
troducing air that has been bacterially filtered.
12
Continue with the "Aeration Procedure" (following).
o
o
C (130 +/- 5
F)
*
. Do not vent sterilizer gas to the

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intellivue trx +

Table of Contents