If you are using ...
6-wire
(Vb is not supported on
TeleMon and is only
supported on MP5 Patient
Monitors with 12-Lead
option)
these leads can be selected at the
Information Center
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5,
V6, V7, V8, V9, V
3R
Sourced (raw) waves are received as:
Channel 1 = II
Channel 2 = III
Channel 3 = Va
Channel 4 = Vb
Defaults are II, Va = V2, III,Vb = V5.
The two chest leads, Va and Vb, can be placed
on the patient in any of the V lead positions (V1
through V9, V3R, V4R, V5R). Lead assignment
is available at the Information Center. When
unassigned, the chest leads use the defaults.
The lead label assigned to Vb cannot be
Note—
selected for Va even though Vb does not appear
to be used.
When display of the pleth wave is
Note—
enabled at the Information Center, the second
chest lead (Vb) is not available for monitoring.
Measuring ECG
, V
, V
.
4R
5R
ECG Monitoring
6-7