Advertisement

Alarm Behavior with Telemetry Overview
Alarm/INOPs
at the
Information
Center
10-6
Pairing with IntelliVue Patient Monitors
Alarm
Source
Bedside and/or
telemetry alarm
Bedside and/or
telemetry alarm
If the transceiver is connected to TeleMon, silencing an active alarm at
Note—
TeleMon silences the alarm at TeleMon only. It has no effect on the alarms at
the Information Center.
The alarms and INOPs that are displayed, recorded, and stored at the
Information Center depend on the type of alarm.
Type of Alarm/INOP
All ECG alarms and INOPs
based on ECG source
Bedside non-ECG alarms and
non-ECG INOPs
Effect at
Where
Paired
Silenced
Bedside
Information
Bedside and/or
Center
telemetry alarm
is silenced
Bedside
Bedside and/or
Silence button
telemetry alarm
is silenced
Effect at Information Center
Displayed, recorded (if
configured), and stored
Displayed, recorded (if
configured), and stored
Effect at
Information
Center
Bedside and/or
telemetry alarm
is silenced
Bedside and/or
telemetry alarm
is silenced (if
Global Silence is
enabled at the
Information
Center)