Cleaning and Sterilization

Cleaning and Sterilization
11-4
Maintenance, Cleaning & Troubleshooting
Step
Action
8
Select Assign Label to initiate programming of the Equipment
Label into the replacement telemetry device
- AND -
within 10 seconds, press the Check button on the telemetry device.
If 10 seconds pass without a button push, then repeat starting
Note—
with step 5.
When the label assignment is complete, the assignment will be
verified in the "Status" field.
9
Go back into Sector Setup, and select the Bed Label and
Equipment Label, followed by OK.
The transceiver goes through the normal startup sequence and starts
monitoring. Verify the signal in the appropriate Patient Sector.
The procedures in this section keep the transceiver clean and provide protection
against infectious agents and bloodborne pathogens. Both the outside of the
transceiver and the inside of the battery compartment must be kept free of dirt,
dust, and debris. The procedures in this section cover the following activities:
Cleaning: removing surface contaminants from the device.
EO Sterilization: using EO gas treatment to decontaminate cleaned
equipment.
After exposure, the transceiver must be cleaned or cleaned and
Important—
EO sterilized as per the instructions contained herein.