Warning
Warning
If the remote Silence key in the Overview window is enabled for IntelliVue
monitors connected to the Information Center, remote silencing for these
beds may be enabled in other clinical units.
Alarm
Source
Bedside alarm
Telemetry
alarm
Bedside and/or
telemetry alarm
Alarm Behavior with Telemetry Overview
Effect at
Where
Paired
Silenced
Bedside
Bedside
Alarm is
Silence button
silenced
Bedside
No effect on
telemetry
alarms
Telemetry Data
Bedside and/or
Window at
telemetry alarm
bedside (if
is silenced
configured)
Pairing with IntelliVue Patient Monitors
Effect at
Information
Center
Bedside alarm is
silenced. There is
no effect on
telemetry alarms
(if Global
Silence is not
enabled at the
Information
Center)
No effect on
telemetry alarms
Bedside and/or
telemetry alarm
is silenced (if
Silence
Overview
Alarms is
enabled at the
Information
Center)
10-5