Paired Device Synchronized Alarm Settings

If the patient's ECG is initially being measured with a patient monitor, and then
the patient is connected to the transceiver for monitoring, the Information Center
will use the patient monitor settings for the transceiver. In general, the
following settings will be synchronized:
Heart Rate
ECG
Arrhythmia
ST
SpO
T
2
When transceivers and MP5 Patient Monitors are paired directly via
Important—
an interface cable, your ECG source must be connected before connecting or
disconnecting the interface cable to maintain synchronized alarm settings. The
interface cable must be disconnected prior to unpairing devices at the
Information Center or the monitor.
Paired Device Synchronized Alarm Settings
HR/Pulse Alarm On/Off, Heart Rate High/Low Limit
Primary Lead, Secondary Lead, Va Lead, Vb Lead
Analysis Mode, Asystole Threshold, Pause
Threshold, VTach HR, VTach Run, PVCs/min, Vent.
Rhythm, SVT HR, SVT Run, PVCs/min On/Off,
Pacer not Capture On/Off, Pacer not Pace On/Off,
Non-sustain On/Off, Vent. Rhythm On/Off, Run
PVCs On/Off, Pair PVCs On/Off, Missed Beat On/
Off, Pause On/Off, R-on-T On/Off, Vent. Bigeminy
On/Off, Vent. Trigeminy On/Off, Multiform PVCs
On/Off, Irregular HR On/Off, SVT On/Off
ST Analysis On/Off, ST Alarm On/Off, ISO point, J
point, ST point, ST Priority List, Single ST Alarm
Limit, Multi ST Alarm Limit
SpO
Alarms On/Off, SpO
2
NBP Alarm Suppression On/Off, Pulse (SpO
Off
Pairing with IntelliVue Patient Monitors
Alarm Limits
2
) On/
2
10-7