Paired Device Synchronized Alarm Settings (Networked) - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Pairing Networked Devices
Paired Device
Synchronized
Alarm
Settings
(Networked)
Heart Rate
ECG
Arrhythmia
ST
QT
SpO
10-14
Pairing Monitoring Devices
Draft - 1 Aug 08
If the patient's ECG is initially being measured with a patient monitor, and
then the patient is connected to the transceiver for monitoring, the Information
Center will use the patient monitor settings for the transceiver. When the
initial ECG source is the transceiver, and then the patient is connected to the
monitor, the Information Center uses its Telemetry Setup settings. The
following settings will be synchronized:
T
2
HR/Pulse Alarm On/Off, Heart Rate High/Low Limit
Primary Lead, Secondary Lead, Va Lead, Vb Lead
Analysis Mode, Asystole Threshold, Pause
Threshold, VTach HR, VTach Run, PVCs/min, Vent.
Rhythm, SVT HR, SVT Run, PVCs/min On/Off,
Pacer not Capture On/Off, Pacer not Pace On/Off,
Non-sustain On/Off, Vent. Rhythm On/Off, Run
PVCs On/Off, Pair PVCs On/Off, Missed Beat On/
Off, Pause On/Off, R-on-T On/Off, Vent. Bigeminy
On/Off, Vent. Trigeminy On/Off, Multiform PVCs
On/Off, Irregular HR On/Off, SVT On/Off, Afib On/
Off
ST Analysis On/Off, ST Alarm On/Off, ISO point, J
point, ST point, ST Priority List, Single ST Alarm
Limit, Multi ST Alarm Limit
QT Analysis On/Off, QTc High On/Off, QTc High
Alarm Limit, dQTc High On/Off, dQTc High Alarm
Limit, QT Lead, QTc Correction Formula, QT
Baseline
SpO
Alarms On/Off, SpO
2
NBP Alarm Suppression On/Off, Pulse (SpO
Off
Alarm Limits
2
) On/
2