Philips IntelliVue TRx Instructions For Use Manual page 101

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Draft - 1 Aug 08
If you are using ...
5-wire (EASI mode)
these leads can be selected at the
Information Center
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5,
V6
In EASI mode, the sourced (raw) waves are
received as:
Channel 1 = Vector 1 (A-I)
Channel 2 = Vector 2 (A-S)
Channel 3 = Vector 3 (E-S)
Defaults are II, V2, III, V5.
Arrhythmia monitoring is performed only
Note—
on the primary and secondary leads selected at
the Information Center, although you can view
and perform ST analysis on all 12 EASI derived
leads.
Measuring ECG
ECG Monitoring
6-7