Pairing Networked Devices
Where Initiated
Bedside
Information Center
Transceiver
(Remote Suspend
enabled at the
Information
Center.)
Bedside
Information Center
Bedside
Information Center
10-10
Pairing Monitoring Devices
Draft - 1 Aug 08
Effect at Bedside
Suspend/Pause
Suspends/Pauses bedside alarms.
Suspends/Pauses bedside and
telemetry alarms.
Suspends/Pauses bedside and
telemetry alarms.
Standby
Bedside put into standby state.
Bedside and telemetry put into
standby.
Resume Monitoring
Bedside resumes monitoring.
Bedside and telemetry resumes
monitoring.
Effect at Information Center
Bed Alarms Suspend/Bed Alarms
Paused INOP displayed.
Suspends/Pauses bedside and
telemetry alarms. Alarms Suspend/
Alarms Paused INOP displayed. All
Arrh Alrms OFF displayed.
Suspends bedside alarms for
Note—
the minimum of the transceiver
configuration and the bedside
configuration.
Suspends/Pauses bedside and
telemetry alarms. Alarms Suspend/
Alarms Paused INOP displayed. All
Arrh Alrms OFF displayed.
Suspends bedside alarms for
Note—
the minimum of the transceiver
configuration and the bedside
configuration.
Monitor Standby INOP displayed.
Monitor and Telemetry Standby
INOPs displayed. Standby location
displayed.
Bedside resumes monitoring.
Bedside and telemetry resumes
monitoring.