Measuring ECG
There is no cardiotach within the transceiver; cardiotach analysis resides in the
arrhythmia algorithm at the Information Center. Therefore, arrhythmia analysis
is always turned on for telemetry patients. Arrhythmia analysis is either Basic or
Enhanced, depending on the product configuration.
ECG Leads
Depending on the leadset connected to the transceiver, a different set of
Monitored
viewable leads are available at the Information Center. The transceiver can
source up to four raw ECG waves.
If you are using ...
3-wire
5-wire (Standard mode)
6-6
ECG Monitoring
Draft - 1 Aug 08
these leads can be selected at the
Information Center
I, II, III
Sourced (raw) waves are received as:
Default is II.
I, II, III, aVR, aVL, aVF, MCL and V
Sourced (raw) waves are received as:
Defaults are II, V, III.
Channel 1 = I, II, or III
Channel 1 = II
Channel 2 = III
Channel 3 = MCL