Philips IntelliVue TRx Instructions For Use Manual page 86

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Technical Alarms (INOPs)
Alarm Text
CENTRAL: TELE ONLY
Source - Patient Monitor
CHECK ECG SETTINGS
Source - Telemetry
CHECK ECG SOURCE
Source - Patient Monitor
CHECK PAIRING
Source - Telemetry
CHECK SpO
T Settings
2
Source - Telemetry
5-12
Alarms
Draft - 1 Aug 08
Technical Alarms (Inoperative Conditions)
Priority
Condition
Hard
System connectivity via the
transceiver is limited (no
alarms, only local
numerics) when in
companion mode and host
monitor does not have
system connectivity. Only
the transceiver's
measurements and alarms
can be displayed at the
Information Center.
Hard
Synchronization of ECG
settings between the
monitor and the
Information Center has
failed.
Yellow
The transceiver and the
Technical
monitor both have valid
Alarm
ECG signals.
Yellow
There is a problem with
Technical
device pairing.
Alarm
Hard
Synchronization of SpO
settings between the
monitor and the
Information Center has
failed.
What to do
For more information only.
Check that the ECG settings
in use are appropriate.
When transitioning
Note—
between networked and non-
networked monitoring , this
INOP will display. Pressing
the Silence button will
dismiss the INOP.
Unpair the transceiver and
monitor or remove the
leadset from the device that
is no longer the ECG source.
Check that the bedside
monitor and telemetry device
are correctly paired.
T
Check that the SpO
2
settings in use are
appropriate.
T
2