Pairing Networked Devices
Pairing with
When the transceiver, assigned to a patient sector at the Information Center, uses
a Direct
an interface cable to connect to an MP5 Patient Monitor, the Information Center
Connection
automatically pairs (if configured to do so) the two devices. Pairing using this
to the MP5
method provides up to 4 ECG waves at the MP5, along with SpO
measurements and a pleth wave. Lead selection and alarm limits may be
changed either at the Information Center or the MP5 Patient Monitor.
To connect the transceiver to the monitor, perform the following steps:
When the transceiver is directly connected to the monitor, arrhythmia
relearning is initiated, and again when the transceiver is disconnected from
the monitor. During relearn, a short dropout period may occur. Be sure to
check your patient's arrhythmia annotation for accuracy whenever relearn
has occurred.
Controls on the transceiver (e.g. nurse call) will be inactive when the device
is directly connected to the monitor except in the case when the monitor has
no network connection and data are transferred via the transceiver.
10-4
Pairing Monitoring Devices
Draft - 1 Aug 08
Step
Action
1
Remove the protective cover from the transceiver's monitor/service
port.
2
Insert the interface cable into the monitor/service port.
3
Attach the interface cable to the monitor using the port labeled with
the
icon.
Warning
Warning
2