Philips IntelliVue TRx Instructions For Use Manual page 37

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Draft - 1 Aug 08
Warning
Warning
If you measure SpO
on a limb that has an inflated NBP cuff, a non-
2
pulsatile SpO
technical alarm can occur. If the monitor is configured to
2
suppress this alarm, there can be a delay of up to 60 seconds in indicating
critical patient status, such as sudden pulse loss or hypoxia.
Warning
Warning
Removal of the SpO
sensor during Continuous SpO
2
a "No Sensor" technical alarm. Silencing this technical alarm turns the
SpO
measurement off. There is no technical alarm for a "No Sensor"
2
condition in Spot Check mode.
Warning
Warning
This equipment is not suitable for use in the presence of a flammable
anesthetic mixture or oxygen concentrations greater than 25% (or partial
pressures greater than 27,5 kPa /206.27 mmHg).
SpO
monitoring results in
2
Product Safety
2-15
2