Philips IntelliVue TRx Instructions For Use Manual page 89

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Alarm Text
NO SIGNAL
Source - Information
Center
OUT OF AREA
Source - Information
Center
REPLACE BATTERY
REPLACE BATTERY T
Note—
This INOP may also
be configured to display as
a Red or Yellow Technical
Alarm.
Source - Telemetry
SOME ECG ALRMS OFF
Source - Information
Center
Draft - 1 Aug 08
Technical Alarms (Inoperative Conditions)
Priority
Condition
Hard
Transceiver is
outside the coverage
area, or
No batteries in
transceiver, or
Transceiver has
failed, or
Connection to
Database Server has
failed.
Short-range radio
connection has
failed.
Hard
The transceiver is
communicating with an
access point designated as a
boundary limit.
Red or
Dead battery. No
Yellow or
monitoring is occurring.
Hard
Technical
Alarm,
Latched
Soft
Some yellow arrhythmia
alarms have been turned off
for this patient.
Technical Alarms (INOPs)
What to do
Make sure that the
transceiver is within
the coverage area and
has good batteries.
Replace the transceiver
if Power On Self Test
fails.
Put bed in Standby.
Contact Service.
For more information only.
Replace batteries.
For information only.