Measuring SpO
2
Continuous
When the transceiver is configured for Continuous SpO
Measurement
"Patient Configurable Settings in Telemetry Setup" on page 9-6), use the
following directions to initiate Continuous SpO
Displaying
When operating in Continuous mode, you can view the Pulse rate at the
Pulse Rate
Information Center. See "Patient Configurable Settings in Telemetry Setup" on
page 9-6.
Note—
Information Center display the parameter label with "T" (for example,
"PulseT") to indicate that the measurement was made via telemetry.
Displaying
The transceiver can be configured to transmit the Pleth wave for display at the
Pleth Wave
Information Center. See "Patient Configurable Settings in Telemetry Setup" on
page 9-6.
8-18
SpO
Monitoring
2
Draft - 1 Aug 08
Step
Action
1
Insert the SpO
sensor light turns on.
2
Attach the sensor to the patient.
3
After approximately 15 seconds, the value, with the measurement
time, is displayed at the Information Center.
There are no sounds associated with continuous SpO
Note—
measurement.
4
To discontinue SpO
the Information Center.
Note—
If the sensor is removed without discontinuing SpO
monitoring in the Telemetry Setup Window at the Information
Center, an SpO
Silencing this technical alarm will turn the SpO
If Pulse rate is turned on, the Patient Sector and Patient Window of the
2
cable into the IntelliVue TRx
2
monitoring, uncheck the Enable SpO
2
T NO SENSOR technical alarm will result.
2
measurement (see
2
monitoring.
+
, and check that the
2
box at
2
2
measurement off.
2