Alarm Text
SpO
T NOISY SIGNAL
2
Source - Telemetry
SpO
T NON-PULSATILE
2
When paired
Note—
directly with an IntelliVue
MP5 Patient Monitor, the
INOP will display as
SpO
T SENSOR OFF.
2
Source - Telemetry
SpO
T SENSOR MALF
2
Source - Telemetry
SRR INTERFERENCE
Source - Patient Monitor
SRR INVALID CHAN
Source - Patient Monitor
SRR MALFUNCTION
Source - Patient Monitor
TAAP DISABLED
Source - Patient Monitor
Draft - 1 Aug 08
Technical Alarms (Inoperative Conditions)
Priority
Condition
Hard
Excessive patient
movements or electrical
interference are causing
irregular pulse patterns
Hard
Pulse is too weak or not
detectable
Hard
Malfunction of the SpO
sensor/adapter cable
Hard
The short-range radio
connection has interference
from another device.
Hard
The channel configuration
of the short-range radio is
invalid.
Hard
There is a malfunction in
the short-range radio
device.
Hard
The monitor's telemetry
configuration does not
allow connection of the
transceiver to the monitor.
Technical Alarms (INOPs)
What to do
Reduce movement or
electrical noise sources.
Check connection to patient.
Change sensor site. Avoid
site distal to BP cuff or intra-
arterial line.
Replace sensor.
2
Move transceiver away from
other interference sources. If
INOP persists, contact
Service.
Check channel and channel
mask configuration.
Contact Service.
If the INOP persists, contact
Service.
Check configuration settings
at the monitor.
Contact Service.
Alarms
5-17