How The Qt Analysis Algorithm Works; Adjusting Qt Settings - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

How the QT Analysis Algorithm Works

How the QT Analysis Algorithm Works
The Information Center measures the QT values once every minute during
startup. Subsequently, the Information Center updates the QT values every five
minutes. Normal or atrial paced beats and beats with a similar morphology are
averaged to form a representative waveform for further processing. Normal
beats followed by a premature QRS will be excluded from the measurements to
prevent the premature beat from obscuring the end of the T-wave. If the
algorithm cannot form a representative waveform, for example because the
morphology of the beats is too varied, the Information Center generates a
Cannot Analyze QT INOP when it detects two consecutive invalid 5 minute
values. This is also the case if normal beats have been falsely labelled so that the
algorithm does not have enough valid beats to make QT measurements. No QT
value is calculated if the QT-HR is >150 bpm (Adult) or >180 bpm (Pedi/Neo).
Because of the different algorithm approaches, a QT/QTc measurement from a
diagnostic 12-lead program may differ from the realtime measurement.
For QT interval monitoring to be effective, basic or enhanced arrhythmia
monitoring must be on.
Adjusting
For patients being monitored by an IntelliVue Telemetry System device you can
QT Settings
adjust QT settings in the QT Setup window at the Information Center. If the
patient is monitored by an IntelliVue Patient Monitor, QT/QTc analysis is
provided by the IntelliVue Patient Monitor. Adjust QT settings at the bedside.
To adjust the settings at the Information Center:
7-18
ST/AR Arrhythmia Monitoring
Draft - 1 Aug 08
Step
Action
1
Access the QT Setup window by selecting the
the All Controls window.
2
Make the adjustments on the QT Setup window. The table below
describes each of the available adjustments.
QT Setup
button in

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intellivue trx+

Table of Contents