Philips IntelliVue TRx Instructions For Use Manual page 82

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Physiologic Alarms
Alarm Text
* NON-SUSTAIN VT
* PACER NOT CAPT
* PACER NOT PACE
* PAIR PVCs
* PAUSE
* PVCs >xxx/MIN
** QTc High
**dQTc High
5-8
Alarms
Draft - 1 Aug 08
Physiologic (Patient) Alarms
Priority
Condition
Yellow
A run of Vs having a ventricular Heart
Rate greater than V-Tach limit but lasting
for less than the V-Tach Run limit.
Yellow
No QRS for 1.75 x the average R-R
interval with Pace Pulse (paced patient
only).
Yellow
No QRS and Pace Pulse for 1.75 x the
average R-R interval (paced patient only).
Yellow
Two consecutive PVCs between non-
PVCs.
Yellow
No QRS detected for x seconds. Choices
of >1.5 to 2.5 seconds.
Yellow
PVCs within one minute exceed by the
PVCs/min limit (xxx).
Yellow
QTc value has exceeded the QTc High
limit for more than 10 minutes
Yellow
dQTc High value has exceeded the dQTc
High Limit for more than 10 minutes.
Source
ST/AR
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Enhanced
Arrhythmia
ST/AR
Basic &
Enhanced
Arrhythmia
ST/AR
Arrhythmia
ST/AR
Arrhythmia

Hide quick links: