Philips IntelliVue TRx Instructions For Use Manual page 103

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Reconstructed
Reconstruction of leads from the sourced wave is defined by the calculations in
Leads
the following table. EASI reconstructed leads are a linear combination of all
three raw EASI leads.
Draft - 1 Aug 08
ECG Lead
5-Wire
3-Wire
Standard
I
I
II
II
(default)
(default)
III
III
(default)
-
MCL
-
aVR
-
aVL
-
aVF
-
V
(default)
Clinical Calculations
6-Wire
in terms of electrodes
I
LA-RA
II
LL-RA
(default)
III
LL-LA
(default)
V-LA, where V=C
aVR
RA-(LA+LL)/2
aVL
LA-(RA+LL)/2
aVF
LL-(LA+RA)/2
V-(RA+LA+LL)/3, where V=C
Va
Va-(RA+LA+LL)/3, where
Va=V2 (default) position
Vb
Vb-(RA+LA+LL)/3, where Vb
=V5 (default) position
Measuring ECG
ECG Monitoring
6-9