Troubleshooting; Basic Troubleshooting - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Troubleshooting

Troubleshooting

Basic Troubleshooting

For problems with...
Important—
such as Access Points and Controllers, are grouped in order to provide coverage
for patients throughout the hospital), there will be a break in monitoring of all
transceivers within the Group(s) being modified. Monitoring will resume when
the new Group Settings are in place. Clinical settings will be unaffected by the
reconfiguration.
11-14
Maintenance, Cleaning & Troubleshooting
Draft - 1 Aug 08
ECG measurement
SpO
measurement
2
Batteries
Nurse call
Electromagnetic
Interference
Short Range Radio
Performance
If changes are made to the Group Settings (how telemetry devices,
see...
"Optimizing ECG Measurement Performance"
on page 6-29.
"Using EASI Leads to Troubleshoot" on page 6-
28.
"Technical Alarms (INOPs)" on page 5-10.
"Optimizing SpO
Measurement Performance"
2
on page 8-22.
"Technical Alarms (INOPs)" on page 5-10.
"Battery Information" on page 4-13.
"Self Test" on page 4-11.
Nurse Call may have been turned off for the
patient. See "Telemetry Button" on page 9-6 for
directions on how to turn it on.
"Reducing Electromagnetic Interference" on
page 12-7.
"Short- Range Radio Error Conditions" on page
10-22

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intellivue trx+

Table of Contents