Unit Settings - Telemetry Setup
Control
Function
SpO
Alarm
Turn SpO
2
Information Center.
SpO
Limits
Increment/decrement SpO
2
High
alarm limit by 1 (in %).
SpO
Limits
Increment/decrement SpO
2
Low
alarm limit by 1 (in %).
3-wire
Set the unit default lead.
5-wire, ECG1
Set the unit default lead.
5-wire, ECG2
Set the unit default lead.
5-wire, ECG3
Set the unit default lead.
5-wire EASI,
Set the unit default lead.
ECG1
5-wire EASI,
Set the unit default lead.
ECG2
alarms on/off at the
2
high
2
low
2
Telemetry Functions at the Information Center & TeleMon
Telemetry Functions at the Information Center
Settings
enable (on)
disable (off)
Limit maximum is 100. Limit
minimum is 51 (adult) or 31
(pediatric). High and low
limit must be at least 1%
apart.
Limit maximum is 99. Limit
minimum is 50 (adult) or 30
(pediatric). High and low
limit must be at least 1%
apart.
I, II, III
I, II, III, MCL, AVR, AVL,
AVF, V
I, II, III, MCL, AVR, AVL,
AVF, V
I, II, III, MCL, AVR, AVL,
AVF, V
I, II, III, AVR, AVL, AVF, V
V
, V
, V
, V
, V
2
3
4
5
6
I, II, III, AVR, AVL, AVF, V
V
, V
, V
, V
, V
2
3
4
5
6
Factory
Default
enable
100 (adult,
pediatric)
90 (adult,
pediatric)
II
II
V
III
,
II
1
,
V
1
2
9-13