Power Management
Using the System
The batteries charge whenever the system is connected to AC power.
CAUTION
The batteries are used to ensure quick system startup and easy movement of the system
only. You cannot perform scans while in battery mode.
CAUTION
When shipping the system, make sure the batteries are inserted upside down to prevent
electrical discharge.
NOTE
When the system is turned on and in Sleep Mode, the batteries power the system.
NOTE
When the battery-charge level reaches a critically low state, the system shuts down
automatically.
NOTE
The batteries may need to be replaced. For information on replacing the batteries, see the
system Service Manual.

Sleep Mode

Sleep Mode uses battery power and allows quick startup of the system. It does not eliminate
power consumption and should be used only for short periods of time, such as when you
transport the system between exam locations. The Sleep Mode control (
) is in the upper left
corner of the control panel, next to the
(On/Off) control. When the system is in Sleep Mode,
it freezes live imaging, pauses any open study, exits any QLAB study, and terminates any
remote connection. Awaken the system to resume normal system operation.
EPIQ 7 User Manual 4535 617 25341