Electromagnetic Compatibility
NOTE
U
is the AC power voltage before application of the test level.
T
NOTE
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Electromagnetic Immunity: Environment Guidance
Immunity Test
Electrostatic discharge
(ESD), IEC 61000-4-2
Electrical fast transient/
burst, IEC 61000-4-4
Surge, IEC 61000-4-5
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines, IEC 61000-4-11
EPIQ 7 User Manual 4535 617 25341
IEC 60601 Test Level
± 6 kV contact, ± 8 kV air
± 2 kV for power supply
lines, ± 1 kV for input/
output lines
± 1 kV differential mode,
±2 kV common mode
Drop out <40 msec
Compliance Level
Same as IEC 60601 test
level
Same as IEC 60601 test
level
Same as IEC 60601 test
level
Same as IEC 60601 test
level
Safety
Electromagnetic
Environment Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If you require
continued operation during
power mains interruptions,
Philips recommends that
the system be powered
from an uninterruptible
power supply or a battery.
75