System Maintenance
Cleaning and Maintaining the System
CAUTION
Turn off power before you remove the air filter. Do not turn on power without the air filter
installed.
CAUTION
Ensure that air filter is dry before installing it. Installing a wet or damp air filter can damage
the system.
1. Locate the air filter handle at the bottom of the left side of the system.
2. Pull the air filter handle straight out, until the filter is completely removed from the system.
3. Inspect the filter. Depending on the condition of the air filter, vacuum or rinse the filter with
water to clean it. If the filter is worn out, or cannot be cleaned, replace it with a spare.
4. To reinstall the filter, slide the filter into place at the bottom of the left side of the system.
292
EPIQ 7 User Manual 4535 617 25341