Transducer Holders and Cable Hangers
1
2

Selecting a Transducer

When the system is turned on, the system initializes the last active transducer. If that
transducer is not connected, the system initializes the transducer that is connected to the
leftmost connector, unless you have defined another transducer as the default transducer. If
you have set a default transducer and it is connected, then the system initializes that
transducer. For information on setting a default transducer, see
and Preset" on page
If multiple transducers are connected when the system is turned on, the system selects the
transducer that was in use when the system shut down. You can select among the connected
transducers during system operation. If the system does not recognize a transducer, it will
switch back to the original transducer. Disconnect and reconnect the unrecognized transducer
or move it to a different receptacle.
You can select or deselect a transducer during live imaging without damaging the transducer or
the system.
1. Touch the transducer name to select it.
2. Touch the clinical application preset that you want to use.
142
Transducer holder
Cable hanger
148.
Selecting a Transducer
"Setting the Default Transducer
EPIQ 7 User Manual 4535 617 25341