Sterilizing Transducers By Immersion - Philips EPIQ 7 User Manual

Ultrasound system
Table of Contents

Advertisement

Transducer Sterilization

Sterilizing Transducers by Immersion

The main difference between sterilization by immersion and disinfection by immersion is the
length of time the transducer is immersed. To determine if your transducer requires
sterilization or high-level disinfection, see
page
257.
1. Clean the transducer and cable according to the procedures in
Connector Cleaning" on page
2. Mix the sterilization solution compatible with your transducer according to label
instructions for solution strength.
3. Immerse the transducer in the sterilization solution as shown in the illustration.
4. Follow the instructions on the sterilization label for the duration of transducer immersion
required for sterilization.
5. Remove the transducer from the sterilization solution after the recommended sterilization
time has elapsed.
6. Using the instructions on the sterilization label, rinse the transducer in sterile water up to
the point of immersion, and then air dry or towel dry with a sterile cloth.
7. Examine the transducer for damage, such as cracks, splitting, fluid leaks, or sharp edges or
projections. If damage is evident, discontinue use of the transducer, and contact your
Philips representative.
EPIQ 7 User Manual 4535 617 25341
"Choosing the Correct Transducer-Care Method" on
258.
Transducer Care
"Transducer, Cable, and
275

Advertisement

Table of Contents
loading

Table of Contents