Per Section 6.4 of the Output Display Standard, measurement precision on the following
quantities is determined by making repeated measurements and stating the standard deviation
as a percentage.
Acoustic Measurement Precision
Quantity
Pr is the underated peak rarefactional pressure
measured in megapascals (MPa).
Wo is the ultrasonic power in milliwatts (mW).
f
is the center frequency in megahertz (MHz)
c
(NEMA UD-2 definition).
PII.3 is the derated spatial-peak pulse intensity
integral in joules per square centimeter (J/cm
Acoustic Measurement Uncertainty
Quantity
Pr is the underated peak rarefactional pressure
measured in megapascals (MPa).
Wo is the ultrasonic power in milliwatts (mW).
f
is the center frequency in megahertz (MHz)
c
(NEMA UD-2 definition).
PII.3 is the derated spatial-peak pulse intensity
integral in joules per square centimeter (J/cm
64
Precision (Percentage Standard
Deviation)
Pr: 5.4%
6.2%
<1%
P11.3: 3.2%
2
).
Measurement Uncertainty
(Percentage, 95% Confidence Value)
Pr: ±11.3%
±10%
±4.7%
PII.3: +18% to -23%
).
2
Biological Safety