Safety
Immunity Test
Power frequency
(50/60 Hz) magnetic field,
IEC 61000-4-8
Conducted RF,
IEC 61000-4-6
Radiated RF, IEC 61000-4-3 3 V/m
CAUTION
If the system is connected to other customer-supplied equipment, such as a local area
network (LAN) or a remote printer, Philips cannot guarantee that the remote equipment will
work correctly in the presence of electromagnetic phenomena.
Although most remote devices comply with their applicable standards for immunity, those
device requirements may not be as stringent as those required for medical equipment. It is the
responsibility of the installer and the user of this remote customer-supplied equipment to
ensure that it functions properly in the electromagnetic environment where the system is
installed. Philips suggests that the installer or the user of such a system consult with experts in
the field of electromagnetic compatibility and safety for guidance to ensure the safe and
effective use of the created system.
76
IEC 60601 Test Level
3 A/m
3 VRMS
150 kHz to 80 MHz
80 MHz to 2.5 GHz
Electromagnetic Compatibility
Compliance Level
Same as IEC 60601 test
level
3 mVRMS
3 V/m
EPIQ 7 User Manual 4535 617 25341
Electromagnetic
Environment Guidance
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
For recommended
separation distances, see
"Recommended Separation
Distance" on page
79.
For recommended
separation distances, see
"Recommended Separation
Distance" on page
79.