Approved Cables For Electromagnetic Compliance - Philips EPIQ 7 User Manual

Ultrasound system
Table of Contents

Advertisement

Safety
Electromagnetic Emissions: Environment Guidance
Emissions Test
Radiated emissions, CISPR 11
Conducted emissions, CISPR 11
Harmonic emissions, IEC 61000-3-2
Voltage fluctuations/flicker
emissions, IEC 61000-3-3

Approved Cables for Electromagnetic Compliance

Cables connected to the system may affect its emissions. Use only the cable types and lengths
listed here.
WARNING
Using cables, transducers, and accessories other than those specified for use with the system
may result in increased emissions from, or decreased immunity of, the system.
Approved Cables
Cable
Adult and pediatric ECG three-lead
patient trunk cable (for AAMI and IEC
leadsets)
Adult ECG leadset (AAMI)
72
Compliance
Class B
However, the use of intracardiac
echocardiography (ICE) or
transesophageal echocardiographic
(TEE) probes results in Class A
Radiated Emissions and should be
used only in a hospital environment.
Group 1 Class B
Class B
Complies
Type
Length
2.7 m (9 ft) with
leadset
Electromagnetic Compatibility
Electromagnetic Environment
Guidance
The system uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The system is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Philips Part Number
453561490131/M1671A
EPIQ 7 User Manual 4535 617 25341