Transducer Disinfection By Immersion - Philips EPIQ 7 User Manual

Ultrasound system
Table of Contents

Advertisement

Transducer Care
NOTE
If you use cleaning wipes, it may be unnecessary to rinse the transducer with water. Always
follow your product label recommendations.
3. Examine the device and cable for damage such as cracks, splitting, sharp edges, or
projections. If damage is evident, discontinue use of the device and contact your Philips
representative.

Transducer Disinfection by Immersion

Methods for disinfecting transesophageal (TEE) and non-TEE transducers differ. Observe the
following warnings and cautions and then see the procedure for your transducer. For non-TEE
transducers, see
transducers, see
Transducers with an AER" on page
WARNING
If a premixed disinfectant is used, be sure to observe the expiration date.
NOTE
All imaging transducers are rated at a minimum IPX7 in accordance with IEC529. This rating
indicates that the device is protected against the effects of immersion. Continuous wave
(nonimaging) transducers are rated IPX1. This rating indicates that the device is protected
against the effects of vertically falling water. For all TEE transducers, the control module is rated
IPX1, while the endoscope is rated IPX7, in accordance with IEC529.
CAUTION
Using non-recommended disinfectants, using incorrect solution strengths, or immersing a
transducer deeper or longer than recommended can damage or discolor the transducer and
voids the transducer warranty.
266
"Disinfecting Non-TEE Transducers by Immersion" on page
"Disinfecting TEE Transducers by Immersion" on page 270
273.
Transducer Disinfection
268. For TEE
or
"Disinfecting TEE
EPIQ 7 User Manual 4535 617 25341

Advertisement

Table of Contents
loading

Table of Contents