Transducer, Cable, and Connector Cleaning

The care method for your transducer determines the appropriate disinfectant for your
transducer. For details about compatible disinfectants, see
"Disinfectants Compatibility" on
page
276.
Transducer, Cable, and Connector Cleaning
These general cleaning (low-level disinfection) instructions are indicated for all transducers and
cables. It is important that you clean the transducer and cable according to the following
procedures.
WARNING
Always use protective eyewear and gloves when cleaning and disinfecting any equipment.
WARNING
If a premixed solution is used, be sure to observe the solution expiration date.
WARNING
Transducers must be cleaned after each use. Cleaning the transducer is an essential step
before effective disinfection or sterilization. Be sure to follow the manufacturer's instructions
when using disinfectants.
CAUTION
Attempting to disinfect a cable or connector by using a method other than the one included
here can damage the device and voids the warranty.
CAUTION
Do not allow sharp objects, such as scalpels or cauterizing knives, to touch transducers or
cables.